AC Option for MTM-EC

Part Code: PCS-007408

Allows users to apply AC electric potentials across the contact via a range of waveforms options, including sine, square, triangular, sawtooth , and more.

  • Waveform Amplifier
  • Required cables for connection to MTM-EC + AC
  • Software license for MTM-EC + AC
Request a Quote
Key FeaturesSpecifications

This module extends the capability of the MTM-EC by enabling alternating current (AC) signals or waveform input. It includes a waveform amplifier and upgraded software features that allow control of a function generator.

The AC add-on makes it possible to apply time-varying voltage signals to the contact, which may be required for simulating electrical conditions in systems such as inverter-fed motors or variable-frequency drives. This allows investigation of how oscillating electrical fields affect the lubricated contact interface.

  • Waveform Amplifier
  • Required cables for connection to MTM-EC + AC
  • Software license for MTM-EC + AC
AC Parameters
Data Acquisition Voltage and current channels sampled at up to 80 M samples per second
Frequency Range 0.03 Hz to 500 kHz
Voltage (Pk / Pk) Low Range: 0 V to 20 V
Measurement accuracy: +/- 20 mV
High Range: 0 V to 30 V
Measurement accuracy: +/- 120 mV
Current (Pk / Pk) Low Range: 0 A to 400 mA
Measurement accuracy: +/- 0.4 mA
High Range: 0 A to 580 mA
Measurement accuracy: +/- 20 mA
Waveform Options Sine; square; triangular; sawtooth; and many more