MTM-CC

Part Code: PCS-007934 / PCS-007936

The PCS Instruments MTM-CC kit is an accessory for the MTM2 instrument that enables controlled environmental testing within the MTM’s test pot. It can be retrofitted to existing MTM instruments or supplied pre-installed with new MTM or MTM-EC units.

  • Lubricant performance testing in inert or reactive environments
  • Evaluation of gaseous lubricity in dry or lubricated conditions
  • Investigating the effect of gaseous additives
  • Studying lubricant behaviour in refrigerant atmospheres
  • (With the optional sealed SLIM kit) Analysis of boundary additive film evolution
Request a Quote
Key FeaturesSpecifications

The PCS Instruments MTM-CC kit is an accessory for the MTM2 instrument that enables controlled environmental testing within the MTM’s test pot. It can be retrofitted to existing MTM instruments or supplied pre-installed with new MTM or MTM-EC units.

The MTM-CC allows users to assess friction and wear performance under controlled atmospheric conditions.

The MTM-CC is offered in three kit configurations.

  • MTM-CC Kit - PCS-007934: Kit of all parts need for the normal MTM
  • MTM-CC Mapper Kit – PCS-007935: Kit of all parts needed for MTM SLIM
  • MTM-CC Bundle – PCS-007936: Kit of all parts needed for a normal MTM and SLIM

Additional Kit

  • MTM-CC Grease Scoop Kit – PCS-007962: Grease scoop, and fittings, plus modification of either the Mapper or normal MTM-CC lid

Brochures & Technical Documents

MTM Brochure

MTM-CC Product Flyer

  • Allows the ambient environment inside the test chamber to be controlled.
  • Supports friction and wear testing in alternative gas atmospheres at ambient pressure.
  • Inlet and outlet ports allow gas flow control through the MTM’s pot.
  • Uses high-performance Kalrez and fluorosilicone seals for maximum compatibility with various gases.
  • Highly flexible ball shaft seal ensures gas isolation without affecting traction measurements.
  • Enables realistic, application-specific testing without altering core MTM functionality.
  • Optional MTM-CC SLIM lid for studying boundary film formation under controlled atmospheric conditions when used in conjunction with MTM SLIM accessory.
  • Optional grease scoop kit allows consistent grease supply to the contact during grease tests.
Specifications of an MTM2 system with the MTM-CC kit fitted
Load Up to 75 N
Contact Pressures Up to 1.25 GPa (standard specimens) (Up to 3.1 GPa with alternative specimens)
Speed ± 4 m/s
Slide/Roll Ratio ± 10,000 %
Temperature Range Ambient to 130 °C (below ambient with optional oil cooler)
Test Sample Volume 30 ml
Pressure Ambient Pressure Only (Maximum 2 millibar above ambient pressure. The MTM-CC is not designed to be a pressurised containment vessel and a small amount of gas leakage will occur during use. Appropriate precautions must be taken if testing dangerous or flammable gases. )